Metrohm 882 Compact IC plus – Anion – MCS User Manual

Page 104

Advertising
background image

6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

94

■■■■■■■■

882 Compact IC plus – Anion – MCS

Problem

Cause

Remedy

Separation column – dead
volume at column head.

Install the separation column in reverse
flow direction (if permitted by the leaflet)
and rinse into a beaker.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 55)
.

Greater rise in the
baseline

Suppressor – reduced
capacity.

Regenerate the suppressor (see Chapter
5.11.3.2, page 82)
.

Precision problems -
significant scatter-
ing of the measured
values

Injection valve – sample
loop.

Check installation of the sample loop (see
Chapter 3.12.1, page 36)
.

Sample – rinsing volume
too low.

Increase rinsing time (see Chapter 5.9, page
76)
.

Injection valve – defective.

Request Metrohm Service.

MCS – vacuum too low.

Check connectors. If they are ok:

Contact the Metrohm Service.

Unexpected change
to the retention
times in the chroma-
tograms

Separation column –
diminished separating effi-
ciency.

Regenerate separation column (see Chap-
ter 5.13.4, page 89)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 55)
.

High pressure pump –
defective.

Request Metrohm Service.