4 Connecting the rinsing the solution outlet capillary

Guide the other end of the outlet capillary labeled with waste
rins.
into a sufficiently large waste container and fasten it there.

5 Connecting the regeneration solution inlet capillary

With a short PEEK pressure screw (6.2744.070), fasten the end of
the inlet capillary labeled with regenerant to the pump tubing
connection of the pump tubing which carries the regeneration
solution.

6 Connecting the regeneration solution outlet capillary

Guide the other end of the outlet capillary labeled with waste
reg.
into a sufficiently large waste container and fasten it there.

3.15

Metrohm CO

2

Suppressor (MCS)

3.15.1

General information on the MCS

The Metrohm CO

2

Suppressor (MCS) removes the CO

2

from the eluent

flow. This reduces the background conductivity, improves the detection
sensitivity and minimizes the injection and carbonate peaks.

CO

2

can reach the eluent flow through the sample itself or arise through

the suppression reaction in the MSM/. The CO

2

peak is effectively mini-

mized through connection of the MCS between the MSM and detector.

The principle of the MCS is based on the gas permeability of the fluoropol-
ymer membrane inside the degassing cell of the MCS. The eluent is guided
through a capillary with a fluoropolymer membrane inside the degassing
cell. A vacuum is generated in the degassing cell by the pump, at the
same time, the pump sucks CO

2

-free air – ambient air is aspired through a

CO

2

adsorption cartridge (26-4), which removes the CO

2

. The pressure

and concentration difference in the degassing cell in comparison to the
inside of the capillary now causes the CO

2

to diffuse out of the eluent

flow.

3.15.2

Connecting MCS

The MCS is connected between the MSM and detector.