Metrohm 882 Compact IC plus – Anion – MCS User Manual

Page 87

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

882 Compact IC plus – Anion – MCS

■■■■■■■■

77

2 Aspirating the sample and measuring time

Aspirate a sample typical for the later application and use a stop
watch to measure the time required by the sample to travel from the
sample vessel to the end of the sample loop.

The time measured corresponds to the "transfer time". The rinsing
time should be at least 3 times the transfer time.

Checking the rinsing time

It is possible to determine whether the rinsing time is adequate via a direct
measurement of the sample carry-over. Proceed as follows:

1 Preparing two samples

Sample A: A typical sample for the application.

Sample sample B: Ultrapure water.

2 Determining "Sample A"

Let "Sample A" pass through the sample path for the duration of the
rinsing time, then inject and measure.

3 Determining "Sample B"

Let "Sample B" pass through the sample path for the duration of the
rinsing time, then inject and measure.

4 Calculating the sample carry-over

The degree of the sample carry-over corresponds to the ratio of the
peak areas of the measurement for sample B to the measurement for
sample A. The lower the ratio, the lower the sample carry-over. This
ratio can be modified by varying the rinsing time – thus allowing the
rinsing time required for the application to be ascertained.

Advertising