Metrohm 882 Compact IC plus – Anion – MCS User Manual
Page 65
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
882 Compact IC plus – Anion – MCS
■■■■■■■■
55
Caution
New separation columns are filled with solution and are sealed on both
sides with stoppers. Before using the column, you need to ensure that
this solution is miscible with the eluents used (observe manufacturer's
data).
You can find the separation columns and guard columns currently availa-
ble from Metrohm in the Metrohm IC Column Program, or in the Internet
in the product area Ion Chromatography. A
test chromatogram and an leaflet are provided along with each column.
You can request detailed information on special IC applications in the cor-
responding "Application Bulletins" or "Application Notes", available
in the Internet at
the Metrohm agent responsible free of charge.
Note
The separation column may only be installed after the initial start-up
(see Chapter 4.1, page 57) of the instrument. Until then, use the cou-
pling (6.2744.040) instead of the guard and separation column.
Connecting and rinsing the separation column
1 Connect the separation column
Caution
When inserting the columns, always ensure that these are cor-
rectly inserted corresponding to the flow direction indicated.
■
Remove stoppers from the separation column.
■
Attach the guard column to the input of the separation column.
OR
Connect the input of the separation column to the outlet capillary
of the guard column, using the PEEK pressure screw (6.2744.070)
included.
OR
If no guard column is used (not recommended), connect the col-
umn input capillary to the input of the separation column, using a
PEEK pressure screw (6.2744.070).