Figure 7, Capillary and cable feed-throughs – Metrohm 882 Compact IC plus – Anion – MCS User Manual

Page 34

Advertising
background image

3.7 Capillary and cable feed-throughs

■■■■■■■■■■■■■■■■■■■■■■

24

■■■■■■■■

882 Compact IC plus – Anion – MCS

1

2

3

1

2

3

6

7

4

5

Figure 7

Capillary and cable feed-throughs

1

Capillary feed-through
For feeding capillaries from the front to the
rear of the instrument.

2

Capillary feed-through
Fro feeding capillaries from the front to the
right side of the instrument.

3

Capillary feed-through
For feeding capillaries from the front to the
left side of the instrument.

4

Luer connector
For connecting a (6.2816.020) syringe. For
manual sample feeding.

5

Capillary feed-through
At the door of the instrument. For feeding
capillaries out of the instrument.

6

Capillary feed-through
At the rear of the instrument. For feeding
capillaries out of the detector chamber.

7

Cable feed-through
At the rear of the instrument. For feeding
the detector cable out of the detector cham-
ber.

Do not feed capillaries through the Luer connectors (7-4). The capillaries
are fastened with PEEK pressure screws (6.2744.070) from inside to the
Luer connector. From outside, liquid can be aspirated or injected with a
syringe.