Once the reading for a well is complete, the pulsing triangle resolves into one of the

following symbols indicating the result of the reading.

Symbol

Description

The bead count for all analytes is
20 or higher.

The bead count for one or more
analytes is below 20 bead
events.

No data were found or an error
occurred while reading the well.

The results are displayed in the Raw Data table and the Bead Map.

Raw Data Table

As the wells are read, the Raw Data table is populated with the results. The table

displays the well location and the well type for all formatted wells. As

Bio-Plex

Manager

MP completes reading each well, the Raw Data table is

populated with the results for that well. The fluorescence intensities of the analytes

you selected are displayed for each well.

Note:

During a reading, the array reader detects all of the analytes in the

sample, including any that you have not selected. However, any analytes that

are not selected in the protocol are not included in the table, even if they are

detected. After a reading you can edit the protocol and change the selection of