7

Maintenance Routines

The maintenance routines keep the MAGPIX instrument in good working order and

operating according to the manufacturer’s specifications. Bio-Rad recommends

that you run the recommended routine at the start of each day to minimize having to

keep track of running the correct routines at the right time. In addition to running the

recommended routine, you can run any of the maintenance routines as necessary.

Calibration and Verification Routine

Description of Routine

Calibration normalizes the operation of the MAGPIX analyzer when it acquires data.

Once calibrated, the values remain until you recalibrate. Verification verifies the

operation of the MAGPIX analyzer.

When to Run this Routine

Bio-Plex

Manager™

MP software adds calibration and verification to the Daily Start

Up routine once a week in accordance with the manufacturer’s recommendation.

Therefore, if you run the Daily Start Up routine daily, you will be running the

Calibration and Verification routine as recommended.

In addition, recalibrate the system if the any of the following is true:

The delta calibration temperature exceeds ±5ºC.

The instrument has been moved.

The instrument undergoes hardware maintenance, for example, a part

replacement.