Calibration failures, Verification failures, Calibration failures verification failures – Bio-Rad Bio-Plex Manager™ MP Software Upgrade User Manual
Page 72
6
| Maintaining the MAGPIX System
70
| Bio-Plex Manager MP Software
Calibration Failures
If the calibration routine fails, Bio-Rad recommends you perform the following steps
in the order shown. Performing the steps in order increases the likelihood that when
you run the calibration routine again, it will succeed.
To correct a calibration failure
1.
Check that the beads are in the correct wells.
2.
Check that the correct calibration control values are being used.
3.
Clean the sample probe.
4.
Check that the waste container is not full.
5.
Run the Adjust Probe Height procedure.
6.
Run the recommended maintenance routine.
7.
Run the Calibration and Verification routine.
If calibration fails a second time, contact Bio-Rad Technical Support.
Verification Failures
If the verification routine fails, Bio-Rad recommends you perform the following steps
in the order shown. Performing the steps in order increases the likelihood that when
you run the verification routine again, it will succeed.
To correct a verification failure
1.
Check that the beads are in the correct wells.
2.
Check that the correct verification and/or calibration control values are being
used.
3.
Clean the sample probe.
4.
Check that the waste container is not full.
5.
Run the Adjust Probe Height procedure.
6.
Run the recommended maintenance routine.