Calibration Failures

If the calibration routine fails, Bio-Rad recommends you perform the following steps

in the order shown. Performing the steps in order increases the likelihood that when

you run the calibration routine again, it will succeed.

To correct a calibration failure

1.

Check that the beads are in the correct wells.

2.

Check that the correct calibration control values are being used.

3.

Clean the sample probe.

4.

Check that the waste container is not full.

5.

Run the Adjust Probe Height procedure.

6.

Run the recommended maintenance routine.

7.

Run the Calibration and Verification routine.

If calibration fails a second time, contact Bio-Rad Technical Support.

Verification Failures

If the verification routine fails, Bio-Rad recommends you perform the following steps

in the order shown. Performing the steps in order increases the likelihood that when

you run the verification routine again, it will succeed.

To correct a verification failure

1.

Check that the beads are in the correct wells.

2.

Check that the correct verification and/or calibration control values are being
used.

3.

Clean the sample probe.

4.

Check that the waste container is not full.

5.

Run the Adjust Probe Height procedure.

6.

Run the recommended maintenance routine.