E (Edge) Mode
The LS-7000 Series measures objects
based on the edges in this mode.
Set the edge numbers to positive values beginning with the top of the measuring head.
The following example shows positive edge numbers and measuring portions.
Bottom
Outer diameter 1
Clearance
Outer diameter 2
Top
1E
2E
3E
4E
5E
6E
Set the edge numbers to negative values beginning with the bottom of the measuring
head. The following example shows negative edge numbers and measuring portions.
Bottom
Outer diameter 1
Clearance
Outer diameter 2
Top
-6E
-5E
-4E
-3E
-2E
-1E
Outer diameter 1: Space between 2E and 3E
(+2E to +3E)
Outer diameter 2: Space between 4E and 5E
(+4E to +5E)
Clearance:
Space between 3E and 4E
(+3E to +4E)
Outer diameter 1: Space between –5E and –4E
(–5E to –4E)
Outer diameter 2: Space between –3E and –2E
(–3E to –2E)
Clearance:
Space between –4E and –3E
(–4E to –3E)