4-43

Chapter 4 Function Settings

4

4

Press the

ENT

key.

The input item is entered and the output setting display appears again.

Offset

An offset value is a desired value that can be added to or subtracted from measured
values. If the offset value is set, it is possible to display the offset value immediately on the
screen by using the auto zero function.

Setting Range
The range can be set between –99.99999 and 99.99999 mm. This range varies with the
display unit or minimum display unit setting. Refer to Display Unit on page 4-57 and
Minimum Display Unit on page 4-58.
Input positive or negative values to increase or decrease the set value.

1

Press the

ENT

key when the output

setting is displayed.

2

Press the

FUNC

key three times.

HI

GO

LO

3

Press the

or

key and set the offset value.

PROG No.

PRINT

AREA

DISP

ESC

G

C

ENT

HOLD

ZERO

FOCUS

A

POSITION

HI

GO

LO

Press these keys
to set the value.

The selectable
part flashes.

Press this key to set the value
to a positive or negative one.

4

Press the

ENT

key.

The input item is entered and the output setting display appears again.

Master Calibration with Automatic Zero (Automatic Offset) Function

The automatic zero function sets measured values to the preset size values (i.e., offset
values) of master workpieces. Refer to Measurement of Outer Diameter and Width (with
Two Measuring Heads for Larger Objects) on page 2-4 for an application example. For
the details of the automatic zero function, refer to Automatic Zero Function on page 3-6.