Measuring lines, 4 measur ing p a rt f eat ur es – HEIDENHAIN ND 100 User Manual User Manual
Page 39
ND 120 QUADRA-CHEK
39
1
.4 Measur
ing P
a
rt
F
eat
ur
es
Measuring lines
A minimum of 2 points are required to measure a line. A maximum of
100 points can be probed and will be processed by a fit algorithm to
define the line.
Press the Line Measurement key. The Probe Line
screen will be displayed. Press the key twice to
measure a series of lines using auto repeat.
Move the stage to position the crosshairs over an end
point of the line and press the ENTER key.
Move the stage to position the crosshairs over the
other end point of the line and press the ENTER key.
Press the FINISH key to complete the measurement.
The line position and angle will be shown and the line
feature will be added to the feature list.
Press the Change soft key to change the line fit
algorithm if desired.
Line fit algorithms include:
LSBF: Fit determined by minimizing the sum of the squared point
deviations from the form fit.
ISO: Fit determined by minimizing the form deviation.
A line is probed on the part
The line position and angle are shown and the line
feature is added to the feature list