Electrical connection, 1 installation, and electr ical connection – HEIDENHAIN ND 100 User Manual User Manual

Page 61

Advertising
background image

ND 120 QUADRA-CHEK

61

2.1 Installation, and electr

ical connection

Electrical connection

Electrical requirements

Environmental conditions

The ND 12x meets standards for normal environmental conditions.

Power connector wiring

L: Line voltage (brown)

N: Neutral (blue)

Protective conductor (ground) terminal (green/yellow)

3-conductor (grounded)

Minimum diameter: 0.75 mm

2

Maximum length: 3 meters

Power input:

AC 100 V ... AC 240 V (–15 % to +10 %)

50 Hz ... 60 Hz (±2 %)

max.54 W

Replacable fuse:

T500 mA / 250 V, 5 mm x 20 mm

Operating temperature: 0° C ... 45° C

Storage temperature:

-20° C ... 70° C

Protection (EN 60529)

IP40, IP54 front panel

Weight:

2.6 kg

Power connector wiring

Warning! Risk of electrical shock

A risk of electrical shock exists if this product is not
properly grounded.

To avoid the hazard, always use a 3-conductor (grounded)
power cord and ensure the ground is properly wired to the
building installation.

Warning! Risk of fire

A risk of fire exists if a power cord not meeting minimum
specifications is used.

To avoid the hazard, always use power cords that meet or
exceed the minimum listed specifications.