Electrical connection, 1 installation, and electr ical connection – HEIDENHAIN ND 100 User Manual User Manual
Page 61
ND 120 QUADRA-CHEK
61
2.1 Installation, and electr
ical connection
Electrical connection
Electrical requirements
Environmental conditions
The ND 12x meets standards for normal environmental conditions.
Power connector wiring
L: Line voltage (brown)
N: Neutral (blue)
Protective conductor (ground) terminal (green/yellow)
3-conductor (grounded)
Minimum diameter: 0.75 mm
2
Maximum length: 3 meters
Power input:
AC 100 V ... AC 240 V (–15 % to +10 %)
50 Hz ... 60 Hz (±2 %)
max.54 W
Replacable fuse:
T500 mA / 250 V, 5 mm x 20 mm
Operating temperature: 0° C ... 45° C
Storage temperature:
-20° C ... 70° C
Protection (EN 60529)
IP40, IP54 front panel
Weight:
2.6 kg
Power connector wiring
Warning! Risk of electrical shock
A risk of electrical shock exists if this product is not
properly grounded.
To avoid the hazard, always use a 3-conductor (grounded)
power cord and ensure the ground is properly wired to the
building installation.
Warning! Risk of fire
A risk of fire exists if a power cord not meeting minimum
specifications is used.
To avoid the hazard, always use power cords that meet or
exceed the minimum listed specifications.