Completion handshake, Input buffer – Quantum Data 802BT User Manual

Page 113

Advertising
background image

802BT/802R Video Test Generator User Guide

103

to wait for all preceding commands to be completed before the commands that follow *WAI

are processed. For example, sending the following command line causes a red rectangle

to be drawn and then overwritten by the colorbar image.

IMGL

COLORBAR;

IMGU

;

IMGE

;

RECT

RED 200 200 0 0 GRAYPAT100

Note that the IMGU command merely requests that the current image be rendered

eventually, not necessarily right away nor in a single attempt. By inserting a *WAI

command after the IMGU, the generator is instructed to render the colorbar image first,

before rendering the red rectangle. This command ensures that the generator draws the

red rectangle on top of the colorbar image.

IMGL

COLORBAR;

IMGU

;

*WAI

;

IMGE

;

RECT

RED 200 200 0 0 GRAYPAT100

The generator will wait until the colorbar image has been completely rendered before it

reads and executes the RECT command.

The

*WAI

command is required by the IEEE-488.2-1992 standard, but also works with

GPIB and RS-232 protocols as well.

Completion handshake

The generator returns a > prompt immediately after an FMTU, IMGU, ALLU, BOOT, INIT,

or SCAL command is received, even if these commands have not finished executing. If the

system controlling the generator must know when the process started by one of these

commands has been completed, then append *OPC? to the command string.

For example, the following command causes the generator to wait until all processes have

been completed before responding with the number 1 and sending the > prompt.

FMTL

VGA_M3;

IMGL

FLAT;

ALLU

;

*OPC

?

The *OPC? command is required by the IEEE-488.2-1992 standard, but also works with

GPIB and RS-232 protocols as well.

Input buffer

Since some commands may take longer to execute, the 802 has an input buffer. This input

buffer is 255 characters long and can be written to by the host controller while the 802 is

busy executing or parsing previous messages. If the input buffer becomes full, the 802 will

hold off the controller until there is room in the buffer. For this reason, a program message

cannot be longer than 255 characters including terminator.