Video Test Generator User Guide

253

CALF

analog video CALibration Factors

Class

System calibration settings

Description

Sets the analog output calibration factors to values other than those set by the generator's

own self-calibration function. Each channel is calibrated using two independent zero

calibration points, and two independent slope calibration points. Interpolation equations for

the different video types are used to set the caldac values that controls the output sync

and voltage levels to the correct values.

TV calibration uses two factors for NTSC output, and two factors for PAL output.

Note: Sending bad calibration factors to the generator via the CALF command can hang

the generator. If the generator hangs it will have to be reinitialized.

Command syntax

CALF zero500R, zero50R, slope1500R, slope1000R, zero500G, zero50G, slope1500G,

slope1000R, zero500B, zero50B, slope1500B, slope1000B, ntsc714, pal700, ntsc_blank,

pal_blank

zero500R

Red calibration factor from 0 (most negative) to 4095 (least negative) at -1.0V.

zero50R

Red calibration factor from 0 (most negative) to 4095 (least negative) at -100mV.

slope1500R

Red calibration factor from 0 (maximum swing) to 4095 (minimum swing) at 2.0V.

slope1000R

Red calibration factor from 0 (maximum swing) to 4095 (minimum swing) at 1.4V.

zero500G, zero50G, slope1500G, slope1000R

Green calibration factors.

zero500B, zero50B, slope1500B, slope1000B

Blue calibration factors.

ntsc714

Factor from 0 (maximum swing) to 4095 (minimum swing) for NTSC output level at

714mV. The ntsc714 and ntsc_blank values are interrelated.

pal700

Factor from 0 (maximum swing) to 4095 (minimum swing) for PAL output level at

700mV.

ntsc_blank

Factor from 0 (least negative) to 4095 (most negative) used to set the zero level of the

NTSC output at 0.0 IRE.