Battery status – Vaisala Hydromet Data Collection Platform User Manual

Page 137

Advertising
background image

Chapter 4 ___________________________________________________________ Troubleshooting

VAISALA______________________________________________________________________ 135

logger output gives the DSI486 version as "A", the communication
module used is the earlier version, DSI486.

Battery Status

You can view battery status information by giving the command
battery without parameters.

Example:

/ > battery
Remaining=100
State=FLOAT_CHARGE
U=6.850195
I=2.685547
Type=PB_BATTERY
Capacity=1.200000
Ext.DC=13.132420
Internal temperature=23.511668

Battery voltage (U) and voltage at the +ExtDC terminal are given as
volts, charging current (I) as milliamperes, and capacity as ampere-
hours. The remaining percentage shows how much energy is left in the
battery.

NOTE

If Program flash checksum returns an error, reload the software. If
reloading the software doesn’t work, contact Vaisala.

NOTE

The battery command applies only to the internal battery of the QML
logger up to QML201B.