Figure 36, Calibration window – Vaisala Hydromet Data Collection Platform User Manual

Page 72

Advertising
background image

User's Guide ______________________________________________________________________

70 ___________________________________________________________________M210784EN-E

0802-145

Figure 36

Calibration Window

The available calibration methods are offset calibration and manual
calibration. In offset calibration, you can change the offset value of the
measurement by entering a new value for the sensor reading, and the
offset value is changed accordingly. Offset calibration may be useful,
for example, if you have a reference sensor providing a measurement
based on which you want to calibrate the sensor connected to your
weather station. In manual calibration, you can change the offset and
gain values directly by entering the new values to be used for the
measurement.

The windows for offset and manual calibration are shown in

Figure 37

on page 73

and

Figure 38 on page 75

, respectively.

NOTE

Sensors with their own measurement interfaces and algorithms, for
example, ceilometers, lightning detectors, or present weather
detectors, and manual and counter-type sensors do not appear in the
sensor list in the Calibration view.