Connecting the power supply – Labconco FlaskScrubber Vantage Series Glassware Washers 4540031 User Manual

Page 19

Advertising
background image

Chapter 3: Getting Started

Product Service 1-800-522-7658

15


If you route the drain piping to a floor drain, refer to the figure below for
placement of the drain tubing and trap.

Connecting the Power Supply

Before continuing with this section, be certain your power supply meets the
requirements specified on the washer data plate and in Electrical Requirements in
Chapter 2: Prerequisites

.

Washers should be hard-wired directly into a junction

box. Refer to the figure below and follow the steps below to connect your washer
to the appropriate power source.

Be certain that you check and conform to all local electrical codes. Do not disturb
any internal wiring in the washer.

If you are connecting to certain legs of a 3-phase power system to obtain single-
phase power, be sure to consult a professional to insure codes/standards are
followed. These professionals can determine the appropriate connections for
either a Wye or Delta connected power source.

The supply circuit must consist of number 12 AWG 3-wire cable or heavier.
Washers must be connected to a dedicated service with a 20 amp circuit breaker
or fuse. Wire must comply with all local electrical codes. Type THW or THWN
wire may be used if it complies.

Drain trap

Advertising