Connecting a cable for monitoring washer operation, Controller set-up – Labconco FlaskScrubber Vantage Series Glassware Washers 4540031 User Manual

Page 21

Advertising
background image

Chapter 3: Getting Started

Product Service 1-800-522-7658

17


5.

Make sure electrical power is turned off before making any connections.

6.

Attach customer-supplied power wires to the brown and blue wires, using

wire connectors. Customer-supplied wires must be clamped to the hole where
they pass through the rear of the junction box, if prescribed by local codes. If
it is necessary to use a bushing with an integral clamp at the inlet to the
junction box, remove the plastic bushing supplied with the washer before
installing the new customer supplied bushing.

7.

Connect the customer-supplied ground wire to the green screw.

8.

Replace the cover on the electrical junction box.

9.

Replace the lower panel and middle panel on the washer.

Connecting a Cable for Monitoring Washer Operation

The operation of the Glassware Washer can be monitored using a computer. See
RS-232 Computer Connection in Chapter 4. If the Glassware Washer will be
installed under a counter, the cable that connects the Glassware Washer to the
computer should be connected to the Glassware Washer prior to placing the
washer in its final location. The Glassware Washer output connector is located on
the back of the washer near the bottom on the left side.

Controller Set-Up

The washer control system has been set at the factory to agree with the
configuration of the washer as it was shipped. The control may be customized or
modified if certain optional accessories are added.

To set up the washer control:

1.

Locate the control panel on the front of the washer.

2.

Press and hold the ▼ button and close the handle of the washer by sliding the door

latch knob to the right. Hold for 5 seconds.

3.

The display will show the SET UP screen.

4.

The > on the left side will highlight TEMP UNITS.

5.

Press ◄ or ► buttons to select either °F or °C which will be used on the display to

set and monitor temperature.

Advertising