4. OPERATING GUIDE - DISCHARGING

The battery must be fully charged before testing starts, except for special tests.

WARNING: Always turn off the AC power switch before

connecting or disconnecting a battery.

NOTE: The test, unless otherwise specified in the

manufacturer's CMM/OMM, shall be conducted at room

ambient temperature of 70°F to 85°F (21°C to 29°C).

A. SWITCH OFF MAINS POWER

Turn off AC on-off/reset power switch.

B. SELECT DISCHARGE MODE

Turn the Mode Select knob to the

“Discharge” setting.

C. TURN DOWN DISCHARGE

CURRENT

Repeatedly turn the Ampere Adjust knob fully

counter-clockwise to set discharge current to

zero. This prevents excess current, which

may be too high for the battery, once the

battery is connected and the discharge test

has begun.

CA-1550 CHARGER / ANALYZER - OPERATING MANUAL V1.2