4. OPERATING GUIDE - DISCHARGING

4.7 TEST COMPLETION AND ANALYSIS

The CA-1550 automatically completes the test with the set parameters. It may

however be required to measure the individual cell voltages for nickel-cadmium

batteries during the test (see section 4.7.2).

I.

WAIT FOR DISCHARGE TEST TO AUTOMATICALLY

COMPLETE

During the test the battery voltage, discharge current, and elapsed

discharge times are displayed. The voltage will continuously decrease while the

current remains constant. The test completes when either the set

discharge time or the low cutoff voltage has been reached. An audible alarm

sounds and the displays freeze at their final values.

J. 1. IF TEST PASSED

The green Accept lamp illuminates. This

indicates the discharge has continued

through the duration of the set discharge time

(the red LED on the timer is off) without the

cutoff voltage being reached. The battery is

usually fit for service, after complete

recharge. See 4.7.1 and 4.7.2 for respective

battery type.

2. IF TEST FAILED

The Reject lamp illuminates if the discharge

has been discontinued because the cutoff

voltage has been reached before the

completion of the set discharge time. The red

LED on the timer remains on, but the time is

frozen at the elapsed time at which the test

failed. The battery is in need of reconditioning or

must be rejected for aircraft service. Refer to the

battery manufacturer's maintenance

manual on how to proceed.

4.7.1 LEAD-ACID BATTERY

If the lead-acid battery has been accepted by the CA-1550 discharge test it has

successfully completed the capacity test.

CA-1550 CHARGER / ANALYZER - OPERATING MANUAL V1.2