J. open the ca-1550, K. calibrate ammeter, L. repeat for additional current settings – Lamar Technologies CA 1550 User Manual

Page 33: 4 shunt verification

Advertising
background image

5. CALIBRATION AND MAINTENANCE

b. If readings differ more than ± 0.2 millivolts

(amperes), continue with steps K through

M.

J. OPEN THE CA-1550

If voltmeter differs more than ± 0.2 millivolts (amperes), open the front panel of

the CA-1550 by removing the two screws on the right side of the panel.

WARNING: Calibration needs to be performed with the unit’s cover

removed as well as both mains power and battery connected. It should only be

performed by trained personnel. If performed incorrectly it could result in

electrical shock leading to injury or death.

K. CALIBRATE AMMETER

The ammeter can be found on the back

side of the front panel. Locate the

ammeter trimpot.

Using a 1/16 inch (2 mm) slot blade

precision type screw driver, adjust the

embedded trimpot located at the rear of

the voltmeter. Adjust until voltmeter

reading matches that of the calibrated

voltmeter (at least within ± 0.2 amperes or

millivolts).

L. REPEAT FOR ADDITIONAL CURRENT SETTINGS

To verify the accuracy over the range of the meter, additional current

settings could be checked. Since the CA-1550 digital meters are linear,

checking two or three current settings are adequate to verify the range. The

easiest way to check an additional setting is to repeat steps H and J but

setting a lower charge current in step H.

5.4 SHUNT VERIFICATION

The CA-1550 internal shunt is calibrated and certified by the shunt manufacturer.

The shunt is a linear resistive device consisting of a heavy brass base and heavy

manganin (copper alloy) resistance. It is not necessary to calibrate the shunt,

however the shunt could be verified with the help of an external calibrated shunt and

a millivoltmeter (see Figure 5-3).

CA-1550 CHARGER / ANALYZER - OPERATING MANUAL V1.2