5. CALIBRATION AND MAINTENANCE

5.3 AMMETER CALIBRATION

To calibrate the CA-1550 ammeter, an external calibrated digital voltmeter, set to

the millivolt range, is required as well as a discharged charged battery. The

millivoltmeter will read the millivolts dropped across the CA-1550 internal shunt

when a charge current is applied. The circuit diagram for the calibration can be seen

in Figure 5-2. The calibration steps are outlined below.

EXTERNAL CALIBRATED MILLIVOLTMETER

Figure 5-2. Circuit diagram for ammeter calibration

A. SWITCH OFF MAINS POWER

Turn off the AC on-off/reset power switch.

CA-1550 CHARGER / ANALYZER - OPERATING MANUAL V1.2