WARNING!
Implanted pacemakers may cause the heart rate meter to count pacemaker pulses
during incidents of cardiac arrest or other arrhythmias. Pacemaker patients should be
carefully observed. Check the patient's pulse; do not rely solely on heart rate meters.
Dedicated pacemaker detection circuitry may not detect all implanted pacemaker
spikes; patient history and physical exam are important in determining the presence of
an implanted pacemaker.
Implanted Pacemakers
The R Series device can be configured to detect pacemaker signals from a patient with an
implanted pacemaker and indicate their presence on the display. When pacer pulses are
detected, the device displays a 5mm, vertical, solid line on the ECG trace.
To disable detection of pacemaker spikes:
1. Press the Param softkey.
2.
Press ECG.
3.
Press Disable Pacer Detect.
To re-enable detection of pacemaker spikes:
1. Press the Param softkey.
2.
Press ECG.
3.
Press Enable Pacer Detect.
5-Lead Monitoring
You can perform 5-lead ECG monitoring with the appropriate ECG patient cable. The 5-lead
cable allows you to monitor the following ECG leads:
•
I, II, III
•
aVR, aVL, aVF
•
V1