Ac charger cpr – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 326
C
HAPTER
14
T
ROUBLESHOOTING
14–8
www.zoll.com
9650-0904-01 Rev. K
AC Charger
CPR
Symptom
Recommended Action
1. The Battery indicator is alternately
illuminating green and yellow.
•
Verify battery is installed.
•
Turn unit ON to identify the fault condition.
•
Replace battery pack with a fully charged battery pack.
•
If problem persists, replace battery pack, unplug device
from AC mains for more than 10 seconds and plug
device back into AC mains.
2. LOW BATTERY message appears
on monitor when unit is plugged
into AC mains.
•
Verify that the AC Power Indicator is illuminated.
•
If not, check AC power cord connection at the wall outlet
and at the rear of the device.
•
Replace battery pack with a fully charged battery pack.
•
Unplug device from AC mains, and plug device back
into AC mains.
•
Verify AC mains is working properly.
3. Neither the Battery, nor AC Power
indicator is illuminated when the
device is plugged into AC mains.
•
Unplug device from AC mains, and plug device back
into AC mains.
•
Verify AC mains is working properly.
CHECK CPR PUCK message is
displayed.
•
No action is required. This is an informational message
to indicate that default calibration values are being used
for the CPR puck. The CPR values will be within the
specification.