ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 52
R Series Service Manual
1-42
9650-0903-01 Rev. J
5
If you are using the Fluke® Biomedical CuftLink Simulator, the volume leak reading should be <10.
6
If you are using the Fluke® Biomedical CuftLink Simulator, the volume leak reading for a failure should be >10.
24.5
On the NIBP simulator, press the START TEST
softkey.
Note: You must press the START TEST
softkey within 30 seconds of closing the
valves on the R Series unit.
After approximately 1 minute, a number appears in the upper middle area of
the NIBP simulator display.
If the simulator:
•
displays a Volume Leak reading <5
5
, then the R Series unit has passed the
test.
•
displays a Volume Leak reading >5
6
, then the R Series unit has failed the
test.
•
displays no Volume Leak reading, but maintains a stable pressure reading
at or above 200 mmHg, then the R Series unit has passed the test; there is
no volume leak.
In addition, the R Series displays the simulator’s pressure reading in the “Cuff
Pressure” field.
After approximately 3 minutes, the valves on the R Series unit open.
o
o
24.6
On the NIBP simulator, press the STOP TEST
softkey.
The NIBP simulator terminates the Volume Leak Test.
o
o
24.7
On the R Series unit, press the EXIT softkey
twice.
The R Series returns to the main NIBP Service Mode screen, then to normal
Monitor mode operation.
Record your results on the Maintenance Test Checklist.
Do this...
Observe this...
Pass/Fail