0 synchronized cardioversion test, Synchronized cardioversion test -23 – ZOLL R Series Monitor Defibrillator Rev J User Manual
Page 33
R Series Service Manual
9650-0903-01 Rev. J
1-
23
12.0 Synchronized Cardioversion Test
Tools Needed: Impulse 4000 or equivalent defibrillator analyzer.
Test Setup:
1. Connect the OneStep cable via the adapter (D.N.I #3010-0378) to the defibrillator analyzer.
2.
Select cardioversion on analyzer. Input 1mV ECG signal at 60 BPM.
Do this...
Observe this...
Pass/Fail
12.1
Press LEAD button to select PADS and Size
X1.
12.2
Press the SYNC softkey on the defibrillator.
Enter synchronized cardioversion timing test
mode on the defibrillator analyzer.
NOTE: Press the SYNC softkey again if the unit
is configured for Remote Sync.
Sync appears on display.
Sync markers display on the monitor. The sync marker appears as a down
arrow over the ECG R-wave peaks on strip chart and display.
o
o
12.3
Select 200J.
o
o
12.4
Press the CHARGE button. When the SHOCK
button lights, press and hold the SHOCK
button.
Defibrillator discharges.
Observe that the R-wave to shock delay (sync delay) is less than 60
milliseconds on the analyzer display.
o
o
Record your results on the Maintenance Test Checklist.