R Series Service Manual
9650-0903-01 Rev. J
2-5
CO2 UNIT ERROR
The EtCO
2
sensor or module has detected a hardware
error.
Check that the sensor is properly plugged in.
Re-insert the sensor. Turn R Series unit off, then
on again to reset. perform a mainstream airway
adapter or module zero, as described in “Zeroing
the Mainstream CAPNOSTAT 5 CO
2
Sensor/
Airway Adapter”.
If the problem persists, contact ZOLL Technical
Support.
CO2 WARM UP
The mainstream sensor is warming up. This may take up
to 5 minutes.
Wait for sensor or module to warm up.
If the message persists more than 5 minutes,
replace the sensor.
CPR FAULT 8
ECG processor not receiving CPR data
Replace analog board.
DATA TRANSFERRED
Transfer done message
DEFIB DISABLED
User prompt issued simultaneously with other faults if defib
is disabled.
Possible configuration problem. Replace high
voltage module. Call ZOLL Technical Support.
DEFIB FAULT 76
PD Defib failure during POST.
Replace PD engine.
DEFIB FAULT 77
PD Defib failure while running.
Replace PD engine.
DEFIB FAULT 78
PD Defib functional safety error while running.
Replace PD engine.
DEFIB FAULT 79
PD module Defib/Pace failure during POST.
Replace PD engine.
DEFIB FAULT 80
Undefined error received from the PD module.
Replace PD engine.
DEFIB FAULT 94
PD module reset 3 times without being requested to.
Replace PD engine.
DEFIB FAULT 95
PD module not communicating.
Replace PD engine.
DEFIB FAULT 96
PD module reported an error on the discharge.
Replace PD engine.
DEFIB MAINT. REQ.
More than 5000 discharges of 200J have occurred.
Maintenance is required.
Replace PD engine.
Error Message
Explanation
User
Advisory
Technical Action