The time since the error occurred is measured in Days:Hours:Minutes:Seconds

Days:Hours:Minutes:Seconds

Days:Hours:Minutes:Seconds

Days:Hours:Minutes:Seconds. It only includes

the time when the device is energized. The error is saved in the memory of the device when it is
switched off. The counter continues when the device is switched back on.

The error symbol is no longer displayed after you read the error record.

6.6.2 Error handling

Types of error message

Description of errors and corrective actions

Type of error

Error

code

Description

Error

E

If an error message is displayed in the Error Records (menu

item B.2.12), the measured value is not correct. The output

current goes to the value set in menu items C.3.5 Error

Handling (Output 1) and C.4.5 Error Handling (Output 2). For

more data about the menu items, refer to

C. Advanced setup

on page 80.

Warning

W

A condition that may not have an immediate effect on device

performance.

Error Message

Error

code

Description

Corrective action

Current output

Current output saturated at

maximum value.

W

The output is at its maximum

output value (20 or 20.5 mA)

because the measured value

is now out of the current

output range.

Fill the tank or remove some

of the product until the level

is back in the configured

range.

Make sure that the 20 mA

setting is correctly set for

your installation. If not,

change the value in menu

item C.3.3 (Output 1) or C.4.3

(Output 2).

Current output saturated at

minimum value.

W

The output is at its minimum

output value (4 or 3.8 mA)

because the measured value

is now out of the current

output range.

Fill the tank or remove some

of the product until the level

is back in the configured

range.

Make sure that the 4 mA

setting is correctly set for

your installation. If not,

change the value in menu

item C.3.2 (Output 1) or C.4.2

(Output 2).

External influences

Temperature is out of range

for NAND Flash

W

The ambient temperature is

outside the given range. This

can cause loss or corruption

of data.

Switch off the device until the

ambient temperature is back

in the given range.

Self-test manager

Self Test has failed

E

The device did not self-test

correctly.

Contact the supplier.