Dielectric constant out of

range

W

The dielectric constant, ε

r

,

calculated by the device is

outside the given range. This

only occurs if the dielectric

constant is automatically

calculated (not set by default).

Check that the probe does not

touch the process connection

(short circuit). Attach the end

of the probe or install it in a

stilling well or bypass

chamber.

Check the programmed probe

length in supervisor mode. Go

to Supervisor > Test >

Supervisor > Test >

Supervisor > Test >

Supervisor > Test >

Information > Probe Length

Information > Probe Length

Information > Probe Length

Information > Probe Length.

If the setting does not agree

with the real value, change

this value in Supervisor

mode. Go to Supervisor >

Supervisor >

Supervisor >

Supervisor >

Advanced Setup > Installation

Advanced Setup > Installation

Advanced Setup > Installation

Advanced Setup > Installation

Setup > Probe Length

Setup > Probe Length

Setup > Probe Length

Setup > Probe Length.

Hardware status

Front end out of order

E

The device's hardware is

defective.

Contact your supplier.

5 volts faulty

E

The device's hardware is

defective.

Contact your supplier.

Time base fault or not

running

E

The device's hardware is

defective.

Contact your supplier.

Electronic offset

W

The device's hardware is

defective.

Contact your supplier.

Time out measurement cycle

E

The device's hardware is

defective.

Contact your supplier.

1 You can change the error code (E) to the warning code (W) in the service menu. For more data, contact your supplier.