Metrohm 743 Rancimat User Manual

Page 115

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

743 Rancimat

■■■■■■■■

107

Delta T

Temperature correction: indication of how much the block temperature
has been increased so that the sample reaches the setpoint temperature.

Gas flow

Gas flow through sample.

Use internal pump

Gas supply with internal pump switched on/off.

Start options

Start delay

Waiting time until starting the measurement.

Start mode

Mode for starting the channels:

per channel: all channels were started individually.

per 2 channels: 2 channels each have been started jointly.

per block: the 4 channels of a block were started jointly.

Stop criteria

Time

Time until the automatic stop of the determination.

Conductivity

Conductivity limit value. The determination was automatically ended upon
reaching this value.

Endpoint

Automatic stop of the determination, as soon as every endpoint (induction
and stability time) has been reached.

Actions at end of determinations...

Shut off heater

The heating was automatically deactivated at the end of the determina-
tion.

Shut off gas flow

The gas flow was automatically deactivated at the end of the determina-
tion.