4 Operation

743 Rancimat

■■■■■■■■

61

Temperature

Setpoint temperature of the sample.

Range

50 ... 220 °C

Delta T

Temperature correction: indication of how much the block temperature
has to be increased so that the sample reaches the setpoint temperature.
This value is determined through measurement of the deviation by the
current temperature of the sample from the heating block temperature
using a calibrated, external temperature sensor (see Chapter 4.4.2, page
50)
.

Note

If no temperature sensor is available for determining Delta T, you can
enter the approximate values listed in the table below, which have been
determined with 5 g silicone oil and a gas flow of 20 L/h (air).

Range

-9.9 ... 9.9 °C