7.5 GLP test set
■■■■■■■■■■■■■■■■■■■■■■
172
■■■■■■■■
743 Rancimat
7.5
GLP test set
External tempera-
ture sensor
Type
Pt100 casing element
⌀1.9 mm, in 4-conductor technology; DIN IEC
751 Class B
Calibration
accuracy
0…90 °C: ±0.02 °C
90…220 °C: ±0.03 °C
Max. deviation
of the meas-
ured tempera-
ture from the
true value (50…
220 °C)
±0.3 °C
Test resistor
10 k
Ω
7.6
RS-232 interface
Plug
Sub-D plug, 9-pin (male)
Basic settings
9600 Baud, 8 Bit, 1 stop bit, no parity, XON/XOFF
7.7
Mains connection
Voltage
2.743.0014: 220…240 V ± 10 %
2.743.0015: 100…120 V ± 10 %
Frequency
50…60 Hz
Power consump-
tion
450 W
Fuse
Diameter 5 mm, length 20 mm
100…120 V: 2.0 ATH (slow-acting)
220…240 V: 4.0 ATH (slow-acting)