4.8 GLP functions

■■■■■■■■■■■■■■■■■■■■■■

154

■■■■■■■■

743 Rancimat

Click on the menu item Tools

GLP test Gas flow... and

confirm with [Next >].
The following dialog window appears:

Enter the Gas flow and confirm with [Next >].

The following dialog window appears:

2 Test all channels

Measure the gas flow sequentially for all channels and enter the
values in the corresponding fields K1…K4 in L/h.

After finishing the measurement, confirm the entries with
[Next >].

The dialog window with the test results appears.