5 recording the temperature, Recording the temperature – Metrohm 743 Rancimat User Manual

Page 48

Advertising
background image

4.2 Instrument and Program Settings

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

743 Rancimat

Displaying the gas flow

Proceed as follows:

1 Open dialog window

Click on the menu item Tools

Gas flow control Display

gas flow....

The following dialog window appears:

Note

This option is only available if no determination is running.

4.2.5

Recording the temperature

The temperature can be recorded at any time for each individual heating
block.

The program writes the recorded measured values into a text file contain-
ing the following values:

t(s)

Time in seconds.

Block [°C]

Temperature measured in the heating block.

Ext. Sensor [°C]

Temperature measured with the external tem-
perature sensor.

The text file is saved under the name U#X-YYMMDD-hhmmss.txt in
the folder Log, # denoting the instrument number (1...4), X denoting the
block (A or B) and YYMMDD-hhmmss denoting the date and time when
the data logging started.