6 "documentation" tab – Metrohm 743 Rancimat User Manual
Page 79

■■■■■■■■■■■■■■■■■■■■■■
4 Operation
743 Rancimat
■■■■■■■■
71
AOCS Cd12b-92
Formula calculation according to standard AOCS Cd12b-92 with Tar-
get temperature = 110 °C.
AOM
Formula calculation according to standard AOM with Target temper-
ature = 97.8 °C.
free
Formula calculation with freely selectable Target temperature.
Formula parameters
Factor
Parameter for the calculation of the standard time, which indicates the
factor by which the time changes per 10 °C. This value can be determined
through the extrapolation (see Chapter 4.7.4, page 117) of results, which
have been recorded at different temperatures.
Range
0.001 ... 100
Target temperature
Target temperature for the calculation of the standard time.
Range
20 ... 300 °C
4.5.2.6
"Documentation" tab
This tab contains parameters for the automatic report output.