3.2 Mounting accessories

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

743 Rancimat

Now fix the air tube onto the reaction vessel cover by firmly pull-
ing the thread adapter M8 / M6.

If determinations are carried out with highly foaming samples,
clamp the foam barrier (5-9) onto the air tube.

Place the reaction vessel cover on the reaction vessel.

Warning

The foam barrier can melt if it projects too deeply into the heating
block.

Ensure that the foam barrier (5-9) is at least 7 cm above the base of
the reaction vessel (5-10).

3.2.4

Inserting vessels / Establishing tubing connections

After you have assembled the reaction and measuring vessels, insert them
in the Rancimat and establish the tubing connections as follows:

1 Insert the measuring vessel

Fill the measuring vessel (5-20) with distilled water.

Place the measuring vessel cover on the measuring vessel.

Insert the measuring vessel into the openings provided on the
Rancimat. At the same time, carefully insert the connector plugs
(5-15) into the electrode connector (3-2).

Connect the white silicon tubing (5-2) to the tubing adapter
M8 / olive (5-11) of the measuring vessel cover.

2 Mount the tubings for the air supply

Screw the FEP tubings 250 mm onto the air supply connections
(3-3) of the Rancimat.

3 Insert the reaction vessel

Fill the reaction vessel (5-10) with the sample.

After reaching the required temperature, insert the reaction vessel
with the mounted reaction vessel cover in the openings provided
on the Rancimat.

4 Establish the tubing connections

Connect the white silicone tubing (5-2), which is fastened on the
measuring vessel cover, to the tubing connector (5-4) of the reac-
tion vessel cover.