HEIDENHAIN IK 5394-3D User Manual

Page 491

Advertising
background image

6 - 41

Quadra-Chek

®

5000

Vector probing increases probing accuracy by modifying the probe path to be-
come orthogonal to the target surface at contact.

Check the Record probed points... box to
enable vector probing.

CAUTION

Some probing applications might include tight part geometry

that doesn’t provide adequate space for vector probing using

default parameters. The probe approach distance can be reduced

in the Probe data path screen of the Probe menu if required. Be

careful to provide enough distance to avoid probe collisions.

NOTE

More detailed information regarding probe path configuration

is included in the discussion of the Probe Path Data screen of

the Probe menu.

With vector probing

Boss surface

Approach distance

Advertising