Chapter 1
_________________________________________________________________________________________

_____________________________________________________________________________

BE6021-11

1316A-2 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 12 of 97

Fig.1.2 Simple NDIR Gas Monitor

1.4.2 The Infrared Transmission Spectra.

The concentration of a subject gas in the sample is computed as a

function of the light intensity measured by the detector. The figure

below characterizes the band pass filters for Freon and CO

2

gases.

Filter

Centre
Wavelength

Band
Width

Centre
Wavelength

Band
Width

μm

μm

cm-1

cm-1

Freon

8.73

1.24

1145

163

Carbon
dioxide

4.425

0.07

2260

36

Fig.1.3 Infrared optical filters for Freon and Carbon Dioxide.

1.4.3 The NDIR Module Building blocks

Infrared Source
A current-regulated infrared (IR) source provides a photon stream.

Light is collimated and directed through the sample cell to the optical

block. The source is modulated at 1 Hz, with a duty cycle of 50%.