5 span calibration – LumaSense Technologies INNOVA 1316A-2 User Manual

Page 67

Advertising
background image

Chapter 4
_________________________________________________________________________________________

_____________________________________________________________________________

BE6021-11

1316A-2 Multi Gas Monitor

LumaSense Technologies A/S

Instruction Manual

Page 67 of 97

20090630_1316

Surplus flow

approx. 200 ml/min

800 ml/min

Multi Gas Monitor - INNOVA 1316

UD 5091

Gas Inlet

Cal. Gas

Fig.4.12 Typical side-stream connection set-up used during calibration

4.3.5 Span Calibration

Turn ON the Multi Gas Monitor. You need to let the Multi Gas Monitor –

INNOVA 1316A-2 warm up for at least 5 minutes. For optimum span

calibration accuracy, let the Multi Gas Monitor –INNOVA 1316A-2 warm

up for at least 30 minutes.

While the Multi Gas Monitor –INNOVA 1316A-2 is warming up, perform

the following steps:
1) If a New NO Sensor is required see the Status window

figure 4.13

.,

install a new NO sensor Order number VM0201A.

2) If the New O

2

Sensor is required see the Status window

figure 4.14

,

install a new O

2

sensor. Order number VM0200A.

Advertising