3.4.1 Pump In Flow.

This fault might indicate one of the following errors. There is an

occlusion in the sample inlet or outlet. A probable course for this

could be occluded (or blocked) tubing, external filter, inlet filter or a

defective pump.

Corrective actions eliminate occlusion.

3.4.2 New NO Sensor

This fault indicate that the Full-scale output of the NO sensor is

≤80% of full scale value when sensor was installed. The NO sensor

must be replaced with a new NO sensor.

Please refer to section

4.3.8

and

5.1.8

when installing a new NO

sensor

3.4.3 New O

2

Sensor

The O

2

sensor output is detected to be below a specified limit during

Zero/O

2

Span Calibration. Replace the O

2

sensor. Please refer to

section

5.1.8

when installing a new O

2

sensor.

3.4.4 IR signal lost

This fault indicate that the sample Cell is contaminated. The sample

cell and windows must be cleaned. Might also display CO

2

and Freon

data invalid in the Data Status section of the status window see

figure 3.10

.

3.4.5 Pump Out Flow

Sample gas path is occluded at outlet during Zero/O

2

span

calibration.
Corrective action: Clear outlet occlusion. Use the pump on command

see

section 5.1.3

and perform a new Zero/O

2

span calibration see

section 4.2.1

.

Might also display a Zero calibration fail error in the Data status of

the status screen.
CO

2

,

Freon

, NO Zero calibration fail and O

2

span calibration fail will be

displayed.