3.4.6 Ambient Temperature range

The detector temperature is out of range. Measurement accuracy

might be affected. Verify that the monitor is operating within the

specified operating temperature range

3.4.7 Pump Low Flow

This fault might indicate one of the following errors. There is an

occlusion in the sample inlet or the pump is defective. A probable

course for this could be occluded tubing, external filter, inlet filter.

Corrective actions eliminate occlusion.


3.4.8 Leak Check


Not included in the 1316A-2.

3.5 Reset.

At any time it is possible to perform a reset to restart the 1316A-2

monitor.

Pull down the Bench Command Menu

Select Soft Reset.

Following window appear.