The following actions are not allowed during the start-up mode:

Zero calibration.

Span calibration.

3.4.2 Normal Operating Mode

When the Start-up time has elapsed, the Multi Gas Monitor 1316A-2

enters Normal Operating Mode.

All actions are allowed, like measurement, calibration or any other

bench command.

If measurement has not been requested for >2 minutes, the Multi

Gas Monitor 1316A-2 leaves Normal mode and enters Standby

mode.

3.4.3 Standby Operating Mode

The 1316A-2 Gas Monitor’s Pump is turned OFF when the Standby

mode is entered.

The 1316A-2 enters standby mode as a result of any of the following:

2 minutes after any POR if measurement has not been

started.

2 minutes after measurement has been stopped.

During Standby mode, the Pump is turned OFF and the following

actions are not allowed:

Zero calibration.

Span calibration.

This explains why measurement should be started before doing any

calibration.

Normally, Standby mode is terminated, and Start-up mode is

entered, when measurement is being started. If the Standby mode

was activated by a sample cell over-temperature condition, the only

way to resume normal operation is via a software reset (bench

command) or via a hardware reset (power off).

Upon leaving Standby mode, the Multi Gas Monitor 1316A-2 turns

ON the pump, attains thermal stability during a 20-second start-up

mode, and resumes Normal mode operation.