Labconco Glassware Washer Service/Technical Manual

20

SECTION III:

Troubleshooting and Alarms


To be effective in properly servicing laboratory equipment, it

is essential to be able to troubleshoot any problem and gain a
clear and concise root to the problem to accurately correct the
problem. This section is split into two sections. The first is

a chart dealing with Basic Troubleshooting for a wide range of
issues. The second section gives detailed steps at finding the

root causes of the Alarms that may be displayed due to Washer
Failures.

Basic Troubleshooting

PROBLEM

CAUSE

CORRECTIVE ACTION

Poor Drying

Dry Temperature too
Low

Program higher dry temperature

Flask, beaker, or
labware with concave

bottoms not loaded
correctly

Tilt these articles when loading so

that water drains off.

Recondensation

If clean labware is left in the

washer, recondensation may occur.

Open latch after completion of cycle
to allow moisture-laden air to

escape.

Heater(s)

disconnected or

malfunctioned

Reconnect or replace heater.

Blower disconnected

or malfunctioned

Reconnect or replace blower.

Dry time too short

Program longer time.

Poor washing
results

Water is not hot

enough

Check Wash 2 water temperature

setting. Must be at least 120°F
(49°C). Recommended temperature is

150°F (66°C).

Improper loading

Check washer loading instructions in
Chapter 4

in this manual.

Detergent too old or

amount wrong

Replace detergent or change quantity

used.

Detergent caked in
cup

Clean cup. Replace with fresh
detergent. Check water inlet

temperature. Must be at least 120°F