Glassware washer/test report – Labconco SteamScrubbe & FlaskScrubbe Glassware Washers 44204 Series User Manual

Page 67

Advertising
background image

61

Glassware Washer/Test Report

MODEL NUMBER

VOLTS/HZ

M.O. NUMBER

SERIAL NUMBER

INSP/TECH/

DATE

QTY/OF

INSP/TESTS No. CONDITIONS

AND

REQUIREMENTS

INST/LCC# RESULTS PASS

Set-Up

1

Connect tap and pure water.

2

Connect

drain.

3

Connect proper electrical power. (Use 115V, 60 Hz or 230V, 50 Hz)

4

Connect current probe to power line.

5

6
7

Close detergent cup door
Connect computer via RS232
Set up washer control
A. Open door
B. Press and hold ▼ button- Simultaneously close and latch door.
C. Press ▼ button to select function. Press ◄ or ► to actuate
function.
1) Set proper voltage
2) Set Temp Units - °C
3) Set DI Pump to Enable- YES
4) Set Light - ON- (window models only)
5) Set Liquid Deter Opt- NO
6) Set Factory Reset – YES

D. Press RUN

Hi-Pot

1

1000V for 1 minute.

Hi Pot Tester

Temperature

1

Position thermocouple near temperature probe.

Test

2

Press and hold ◄ button - simultaneously close and latch door.

3

Display must agree with temperature meter ±2°C.

Temperature Meter

Self Test

1 Open

door

2

Press & hold ▲ button – Simultaneously close and latch door.

3 Press

▼ button to select function. Press ◄ or ► to actuate function.

TAP

WATER

Visual

PURE

WATER

Visual

Continue in this mode until Float Switch shuts off water flow.

Visual

Open door to check water level. Bottom of heater must be
partially wetted. Top of heater must be partially unwetted.
Close door and resume DIAGNOSTICS.

WASH

PUMP

Visual

Advertising