Alarms – Labconco SteamScrubbe & FlaskScrubbe Glassware Washers 44204 Series User Manual

Page 32

Advertising
background image

Labconco Glassware Washer Service/Technical Manual

26

Low Temp Alarm

Heater inoperative

Check electrical connections.

Check voltage to heater.


Check resistance of heater. If open
circuit, replace heater.

Defective controller Check output from PCB to heater.

Replace PCB. Contact Labconco
Product Service.

Defective

temperature probe

Check electrical connections.


Contact Labconco Product Service.
Replace if necessary.

Defective high

temperature cut out

switch

Check resistance of safety cut out

switch. If open circuit at room
temperature, replace switch.

Alarms

Throughout a wash cycle there are several abnormal events that could occur.
The washer is programmed to monitor these occurrences and provide an alarm to

alert the operator. Information is shown on the display to identify the alarm

situation. The following messages may be displayed. Certain alarm situations

are signaled with an audible sound which self cancels after 1 minute.


HOT GLASS

This displays at the end of a cycle if the temperature inside the

washer is over 122°F (50°C). The alarm cancels when the temperature

cools.

PLEASE WAIT

This displays if the door is opened and then closed during a wash or

rinse segment of a cycle. The display will also show the time remaining

until the operation of the washer will resume where it was interrupted.


LOW WATER

This displays if a water fill is inadequate. The washer will cancel the

cycle and drain water that is in the washer. The alarm is canceled by
opening the door.

WATER HIGH (OVERFILL)

This displays if the washer overfills with water. Water fill valves

will be closed and the washer will cancel the cycle and drain water
that is in the washer. The alarm is canceled by opening the door.


OVERFLOW

This displays if water is detected in the base of the washer. Water

fill valves will be closed and the washer will cancel the cycle and

Advertising