Labconco SteamScrubbe & FlaskScrubbe Glassware Washers 44204 Series User Manual

Page 60

Advertising
background image

Labconco Glassware Washer Service/Technical Manual

54


6.

Through the front, remove the two nuts on the front side of the blower

mounting bracket.

7.

Through the back, remove the two nuts on the back side of the blower

mounting bracket.

8.

Pull the blower out through either the back or front (whatever is

easier).

9.

Disconnect the blower’s wiring from the base wiring harness.

10. Connect the new blower’s wiring to the base wiring harness.

11. Seat the new blower’s bracket over the mounting screws and tighten it

with the nuts removed earlier.

12. Attach the Flex Ducting from the Side Cabinet to the blower’s inlet

collar and tighten the clamp over the fitting.

13. Attach the Flex Ducting from the drying manifold to the blower’s

exhaust collar and tighten the clamp.

14. Test the blower in diagnostics.

Drain System Replacement/Installation

Use the following procedures to replace/install different parts of the
Drain System. IF THE MACHINE IS DISPLAYING “DRAIN FAIL ALARM” FIRST

READ AND GO THROUGH THE DRAIN FAIL ALARM TROUBLESHOOTING CHAPTER. The
parts listed below do not include every part critical to the drain
system; however, those not listed will be mentioned here for reference

in their own sections.

**Completely drain and disconnect electrical power to the unit before

proceeding with any of these instructions.

Replacing the Drain Valve
The drain valve is regulated by the PC Board and allows the water to leave

the tank when the unit is set to drain. The Plastic Solenoid housing for the
115V valve is RED; the 230V valve is GOLD/ORANGE. If it is discovered that

Figure 28
Left:
FlaskScrubber®
Vantage® Blower. Just
like the Blower in the
FlaskScrubber® the
blower (indicated by the
Yellow Bracket) is held to
the base weldment by a
mounting Bracket,
Magenta Bracket.

The Green Arrow shows
the clamp that holds the
flexible ducting that leads
to the manifold, to the
Blower.

The Red Arrow points to
the clamp that secures the
Flex Ducting that comes
from the HEPA filter in
the left Side Cabinet to
the Vantage's® Blower.

Advertising