BPS / MX / RS Series SCPI Programming Manual

AMETEK Programmable Power

Manual P/N 7003-961 Rev. AA

101

PONSetup:CURRent

This command sets the initial current limit at the power on. The maximum allowable current
is controlled by the voltage range and voltage mode.

Command Syntax

[SOURce:]PONSetup:CURRent <NRf+>

Parameters

range set by the LIMit:CURRent, voltage mode and voltage range

Examples PONS:CURR

22

Query Syntax

PONS:CURR?

Returned Parameters

<NR2>

PONSetup:FREQuency

This command sets the initial frequency of the power source at the power on.

Command Syntax

[SOURce:]PONSetup:FREQuncy <NRf+>

Parameters

set by the LIMit:FREQuency

Examples PONS:FREQ

50

Query Syntax

PONS:FREQ?

Returned Parameters

<NR2>

PONSetup:NOUTput

This command sets the configuration of the power source to one phase or three phase if this
option is enabled. The LAST setting causes the unit to power up in the mode it was left.

Command Syntax

[SOURce:]PONSetup:NOUTput <selection>

Parameters ONEPhase|THReephase|LAST
Examples PONS:NOUT

LAST

Query Syntax

PONS:NOUT?

Returned Parameters

<CRD>

PONSetup:OLOad[:MODE]

This command determines how the power source will react to an over current condition. In
const current mode, the output voltage is reduced to a value that maintains a constant laod
current. In constant voltage mode, the output faults to an initial value if the over current
condition is maintained beyond the protection delay period.

Command Syntax

[SOURce:]PONSetup:OLOad[:MODE] <mode>

Parameters CCURrent|CVOLyage
Examples OLOad

CCUR

Query Syntax

PONS:OLO?

Returned Parameters

<CRD>

Related Commands

CURR:PROT:STAT