AMETEK BPS Series Programming Manual User Manual
Page 129

BPS / MX / RS Series SCPI Programming Manual
AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA
129
TRIGger:ACQuire:SOURce
[3Pi Controller Only]
These commands select the trigger source for a triggered measurement sequence as
follows:
IMMediate
Acquire the measurements immediate. The FETCh commands return the
requested calculation from the set of acquired data as soon as the
measurement calculations are completed by the controller.
SYNChronize Internal phase A angle
BUS
IEEE-488 device, *TRG, or <GET> (Group Execute Trigger)
TTLTrg
The signal driving the Trigger Out BNC
Command Syntax
TRIGger:ACQuire:SOURce<CRD>
Parameters IMM|SYNC|BUS
|TTLT
*RST Value
IMM
Examples TRIG:ACQ:SOUR
BUS
Query Syntax
TRIGger:ACQuire:SOURce?
Returned Parameters
<CRD>
Related Commands
ABOR TRIG:SOUR
TRIG:SYNC:SOUR
TRIG:SYNC:PHAS
INIT
INIT:CONT
*TRG
TRIGger:SYNChronize:SOURce
This command selects the synchronizing trigger source in generating a step, pulse, or list
output as follows:
• IMMediate starts the transient output immediately.
• PHASe
starts the transient output at the reference phase set by TRIG:PHAS.
Command Syntax
TRIGger:SYNChronize:SOURce <CRD>
Parameters IMMediate|PHASe
*RST Value
IMM
Examples
TRIG:SYNC:SOUR IMM
TRIG:SOUR PHAS
Query Syntax
TRIGger:SYNC:SOURce?
Returned Parameters
<CRD>
Related Commands
ABOR
TRIG:SYNC:PHAS
INIT
INIT:CONT
*TRG
TRIGger:SYNChronize:PHASe
This command sets the phase angle with respect to an internal phase reference if
TRIGger:SOURce PHASe is selected. The range is from -360 to +360 Degrees.
Command Syntax
TRIGger:SYNChronize:PHASe<NRf+>
Parameters
360 to +360
*RST Value
0
Examples TRIG:SYNC:PHAS
90
Query Syntax
TRIGger:SYNChronize:PHASe?
Returned Parameters
<NR2>
Related Commands
ABOR TRIG:SYNC:SOUR
INIT
INIT:CONT *TRG