AMETEK BPS Series Programming Manual User Manual

Page 187

Advertising
background image

BPS / MX / RS Series SCPI Programming Manual

AMETEK Programmable Power

Manual P/N 7003-961 Rev. AA

187

IEC413:IHARmonics:PAUSe

This command will set the pause time in seconds between each interharmonics frequency.
The class must be selected prior to this command. Query with min and max to find the
range of the pause time.

Command Syntax

IEC413: IHARmonics:PAUSe <NRf+>

Parameters <seconds>
Examples IEC413:IHAR:PAUS

2

Query Syntax

IEC413:IHAR:PAUS?

Returned Parameters

<NR1>

Advertising