BPS / MX / RS Series SCPI Programming Manual
AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA
103
PONSetup:VOLTage:MODE
This command sets the initial voltage mode at the power on. This mode is DC or AC
Command Syntax
[SOURce:]PONSetup: VOLTage:MODE
Parameters
AC | DC | ACDC
[ACDC on 3Pi Controller Only]
Examples
PONS: VOLTage:MODE DC
Query Syntax
PONS: VOLTage:MODE?
Returned Parameters
<CRD>
PONSetup:VRANge
This command sets the initial voltage range of the power source.
Command Syntax
[SOURce:]VRANge<NRf+>
Parameters 150
⎪200 | 300 | 400
Examples PONS:VRAN
150
Query Syntax
PONS:VRAN?
Returned Parameters
<NR2>
PONSetup:WGRoup
[3Pi Controller Only]
This command determines the source of its waveform storage. It has a total of 4 groups.
Change of group requires power down, an IEEE-488 Device Clear (not available when usign
the RS232C interface) or a *RST command to take effect.
Command Syntax
[SOURce:]PONSetup:WGRoup <n>
Parameters 0|1|2|3
Examples PONSetup:WGRoup
2
Query Syntax
PONSetup:WGR?
Returned Parameters
<NR1>
Related Commands
TRACe:CAT?